Cat No: KI27603-INGAAS
An Optical Loss Test Set used to test power, insertion loss, continuity and faults on multimode fiber optic systems at 850 / 1300 nm. In a small package, it provides a powerful solution to improve test flexibility and productivity.
The 650 nm VFL Visitester feature mixes a VFL in with the test signal during Autotest. In addition to operating as a typical VFL, it gives positive visual identification of a fiber to be connected for loss testing.
| Attribute Name | Attribute Value |
|---|---|
| Detector type (meter) | InGaAs |
| Response wavelength (meter) | 600 ~ 1700 nm |
| Calibration wavelengths (meter) | 780, 820, 850, 980, 1270, 1290, 1300, 1310, 1330, 1350, 1370, 1390, 1410, 1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570, 1590, 1610, 1625, 1650 nm |
| Power range (meter) | +10 ~ -60 dBm (780 ~ 980 nm) / +10 ~ -70 dBm (1270 ~ 1650 nm) |
| Mid-range linearity (meter) | 0.04 dB (typical) |
| Total uncertainty (meter) | 0.3dB (max) |
| ID tone detection & Autotest sensitivity (meter) | Up to -50 dBm |
| Modulated output (source) | 270 Hz, 1 kHz, 2 kHz ± 2 %, 12 Multi-Fiber ID tones |
| Output wavelength (source) | Port 3: 850-300 nm LED, VisiTester |
| Output power (source) | -23 dBm ±1 dB (@ 62.5/125 µm for 850, 1300 nm) |
| Adjust output power (source) | Fixed (for 850, 1300 nm) / Adjustable over 7 dB in 0.01 dB steps (for VisiTester) |
| Short term stability for 15 min, typical ±2 °C, after warm up, ORL < -25 dB (source) | 0.01 dB (for 850, 1300 nm) |
| Stability over temperature (source) | 0.35 dB (typical for 850, 1300 nm) |
| Wavelength drift (source) | 0.4 nm per °C (typical for 850, 1300 nm) |
| Mode controlled (source) | 850, 1300 nm (50/125µm compliant: IEC 61280-4-1 {Ed.1.0}, TIA 526-14A & TIA TSB-178) |
| Reconnection repeatability @ 95% confidence (source) | 0.05 dB (for 850, 1300 nm), 0.1 dB (for VisiTester) |
| VisiTester wavelength | 650 ± 5 nm |
| VisiTester spectrum width @FWHM | 3 nm |
| VisiTester wavelength drift | 0.1 nm per °C |
| VisiTester power | +2 dBm ± 1 dB @ MMF |
| VisiTester laser safety | Class 1, IEC60825-2 |
| VisiTester blink rate (manual) | CW or 2 Hz |
| Accuracy (meter) | 1 % (0.06 dB) |
| Calibration validity (ISO/IEC 17025) | 3 years |
| Resolution (meter) | 0.01 dB |
| Polarization sensitivity (meter) | < 0.05 dB (typical) |
| Autotest | 1-way: compatible with Kingfisher's 1-way Loss Test Sets & Power Meters (except KI9600 series) |
| Tone detection (meter) | 150 ~ 9900 Hz ± 1 % |
| Battery life | Laser/LED source: 90/80 hours (typical) in Autotest |
| Power | Battery / External power via USB |
| Battery type | 2x AA (LR6) Alkaline / 2x AA NiMH AA |
| USB interfaces | USB-micro for power & software / USB-A for memory key |
| Auto off | Selectable |
| Software | KITS (free download) |
| Memory | 1,000 four λ tests with date & time / Unlimited on USB |
| Case | Polycarbonate (body) / rubber (edges & corners), moisture resistance, 1-meter drop tested |
| Display size | 74 mm x 55 mm |
| Size | 190 mm x 105 mm x 35 mm |
| Unit weight | 0.42 kg |
| Shipping weight | 1.5 Kg |
| Operating / storage temperature | -15 to 55 °C / -25 to 70 °C |
| Relative humidity | 0 ~ 95 % |
| Warranty | 3 years |